AТ2533, AT2533/1 Dosimeters
Detector Silicon semiconductor detector;
Geiger-Muller counter tube
Energy range 50 keV – 10 MeV
Measurement range of ambient dose equivalent rate Ḣ*(10) 1 µSv/h – 1000 Sv/h
Measurement range of ambient dose equivalent Н*(10) 10 µSv – 5000 Sv
Limits of variation of the response due to dose Н*(10) and dose rate Н*(10) ±15%
Measurement range of average pulsed radiation dose rate 30 µSv/s – 0.3 Sv/s
(100 mSv/h – 1000 Sv/h)
(pulse repetition rate is not less than 20 cps, duration not less than 1 μs)
Measurement range of pulsed radiation dose 10 µSv – 5000 Sv
Limits of variation of the response due to dose and average dose rate of pulsed radiation ±20%
Limits of variation of relative response due to gamma radiation energy (50 keV to 3 MeV) and angle of incidence (0° to ±45°) -29% to +67%
Typical sensitivity to 137Cs gamma radiation 0.15 cps/(µSv·h-1) (for Ḣ*(10) ≤0.1 Sv/h)
58 mV/(Sv·h-1) (for Ḣ*(10) >0.1 Sv/h)
Response time for 10-fold dose rate change ≤10 s (for Ḣ*(10) >10 µSv/h)
Burn-up life ≥25000 Sv (BDKG-33 and cable)
Protection class (BDKG-33) IР68
(withstands static hydraulic pressure up to 400 kPa at 40 m immersion depth)
PC interface USB 2.0 (АТ2533)
USB 2.0 / Bluetooth (АТ2533/1)
Overall dimensions, weight Ø30х130 mm, 0.25 kg (BDKG-33)
85х155х35 mm, 0.3 kg (PU-33)

Design and specifications are subject to change without notice.
For details please refer to the datasheet.

Downloads

You have not viewed any product yet!